Show simple item record

FieldValueLanguage
dc.contributor.authorBegg, J
dc.contributor.authorJelen, U
dc.contributor.authorKeall, P
dc.contributor.authorLiney, G
dc.contributor.authorHolloway, L
dc.date.accessioned2023-01-18T00:25:00Z
dc.date.available2023-01-18T00:25:00Z
dc.date.issued2022en
dc.identifier.urihttps://hdl.handle.net/2123/29895
dc.description.abstractObjective. Reference dosimetry on an MRI-linac requires a chamber specific magnetic field correction factor,kB⃗.This work aims to measure the correction factor for a parallel plate chamber on a parallel MRI-linac. Approach. kB⃗is defined as the ratio of the absorbed dose to water calibration coefficient in the presence of the magnetic field,ND, wB⃗ relative to that under 0 T conditions,ND, w0T. kB⃗ was measured via aND, wtransfer to a field chamber at each magnetic field strength from a chamber with known ND, w and kB⃗. This was achieved on the parallel MRI-linac by moving the measurement set-up between a high magnetic field strength region at the MRI-isocentre and a low magnetic field strength region at the end of the bore whilst maintaining consistent set-up and scatter conditions. Three PTW 34001 Roos chambers were investigated as well as a PTW 30013 Farmer used to validate methodology.Main Results.The beam quality used for the measurements of kB⃗ was TPR20/10 = 0.632. The kB⃗for the PTW Farmer chamber at 1 T on a parallel MRI-linac was 0.993 ± 0.013 (k = 1). The average kB⃗ factor measured for the three Roos chambers on a 1 T parallel MRI-linac was 0.999 ± 0.014 (k = 1).Significance.The results presented are the first measurements of kB⃗ for a Roos chamber on a parallel MRI-linac. The Roos chamber results demonstrate the potential for the chamber as a reference dosimeter in parallel MRI-linacs.en
dc.language.isoenen
dc.publisherIOP Sciencesen
dc.relation.ispartofPhysics in Medicine & Biologyen
dc.rightsCreative Commons Attribution-NonCommercial-NoDerivatives 4.0en
dc.subjectMRI-linacsen
dc.subjectMRgRTen
dc.subjectRoos chamberen
dc.subjectabsorbed doseen
dc.subjectmagnetic fielden
dc.titleExperimental characterisation of the magnetic field correction factor,k->B,for Roos chambers in a parallel MRI-linac.en
dc.typeArticleen
dc.subject.asrc0299 Other Physical Sciencesen
dc.identifier.doi10.1088/1361-6560/ac66b8
dc.type.pubtypeAuthor accepted manuscripten
dc.relation.nhmrc1132471
dc.relation.nhmrc1194004
usyd.facultySeS faculties schools::Faculty of Medicine and Healthen
usyd.departmentACRF Image X Instituteen
usyd.citation.volume67en
usyd.citation.issue9en
usyd.citation.spage095017en
workflow.metadata.onlyNoen


Show simple item record

Associated file/s

Associated collections

Show simple item record

There are no previous versions of the item available.