dc.description.abstractPURPOSE: Until now, real-time image guided adaptive radiation therapy (IGART) has been the domain of dedicated cancer radiotherapy systems. The purpose of this study was to clinically implement and investigate real-time IGART using a standard linear accelerator. MATERIALS/METHODS: We developed and implemented two real-time technologies for standard linear accelerators: (1) Kilovoltage Intrafraction Monitoring (KIM) that finds the target and (2) multileaf collimator (MLC) tracking that aligns the radiation beam to the target. Eight prostate SABR patients were treated with this real-time IGART technology. The feasibility, geometric accuracy and the dosimetric fidelity were measured. RESULTS: Thirty-nine out of forty fractions with real-time IGART were successful (95% confidence interval 87-100%). The geometric accuracy of the KIM system was -0.1 ± 0.4, 0.2 ± 0.2 and -0.1 ± 0.6 mm in the LR, SI and AP directions, respectively. The dose reconstruction showed that real-time IGART more closely reproduced the planned dose than that without IGART. For the largest motion fraction, with real-time IGART 100% of the CTV received the prescribed dose; without real-time IGART only 95% of the CTV would have received the prescribed dose. CONCLUSION: The clinical implementation of real-time image-guided adaptive radiotherapy on a standard linear accelerator using KIM and MLC tracking is feasible. This achievement paves the way for real-time IGART to be a mainstream treatment option.en
dc.publisherElsevieren
dc.relationNHMRC 1112096en
dc.rightsOther
dc.subjectImage-guided radiotherapyen
dc.subjectreal-time imagingen
dc.titleThe first clinical implementation of real-time image-guided adaptive radiotherapy using a standard linear accelerator.en
dc.typeArticleen
dc.subject.asrcFoR::029903 - Medical Physicsen
dc.identifier.doi10.1016/j.radonc.2018.01.001
dc.type.pubtypeAuthor accepted manuscripten
usyd.facultyFaculty of Medicine and Health, Sydney Medical Schoolen