Show simple item record

FieldValueLanguage
dc.contributor.authorOborn, B
dc.contributor.authorGe, Y
dc.contributor.authorHardcastle, N
dc.contributor.authorMetcalfe, P
dc.contributor.authorKeall, P
dc.date.accessioned2018-03-23
dc.date.available2018-03-23
dc.date.issued2016-01-01
dc.identifier.citationMed Phys. 2016 Jan;43(1):368en_AU
dc.identifier.urihttp://hdl.handle.net/2123/18027
dc.description.abstractPURPOSE: To report on significant dose enhancement effects caused by magnetic fields aligned parallel to 6 MV photon beam radiotherapy of small lung tumors. Findings are applicable to future inline MRI-guided radiotherapy systems. METHODS: A total of eight clinical lung tumor cases were recalculated using Monte Carlo methods, and external magnetic fields of 0.5, 1.0, and 3 T were included to observe the impact on dose to the planning target volume (PTV) and gross tumor volume (GTV). Three plans were 6 MV 3D-CRT plans while 6 were 6 MV IMRT. The GTV's ranged from 0.8 to 16 cm(3), while the PTV's ranged from 1 to 59 cm(3). In addition, the dose changes in a 30 cm diameter cylindrical water phantom were investigated for small beams. The central 20 cm of this phantom contained either water or lung density insert. RESULTS: For single beams, an inline magnetic field of 1 T has a small impact in lung dose distributions by reducing the lateral scatter of secondary electrons, resulting in a small dose increase along the beam. Superposition of multiple small beams leads to significant dose enhancements. Clinically, this process occurs in the lung tissue typically surrounding the GTV, resulting in increases to the D98% (PTV). Two isolated tumors with very small PTVs (3 and 6 cm(3)) showed increases in D98% of 23% and 22%. Larger PTVs of 13, 26, and 59 cm(3) had increases of 9%, 6%, and 4%, describing a natural fall-off in enhancement with increasing PTV size. However, three PTVs bounded to the lung wall showed no significant increase, due to lack of dose enhancement in the denser PTV volume. In general, at 0.5 T, the GTV mean dose enhancement is around 60% lower than that at 1 T, while at 3 T, it is 5%-60% higher than 1 T. CONCLUSIONS: Monte Carlo methods have described significant and predictable dose enhancement effects in small lung tumor plans for 6 MV radiotherapy when an external inline magnetic field is included. Results of this study indicate that future clinical inline MRI-guided radiotherapy systems will be able to deliver a dosimetrically superior treatment to small (PTV < 15 cm(3)), isolated lung tumors over non-MRI-Linac systems. This increased efficacy coincides with the reimbursement in the United States of lung CT screening and the likely rapid growth in the number of patients with small lung tumors to be treated with radiotherapy.en_AU
dc.publisherAmerican Institute of Physicsen_AU
dc.relationNHMRC 1036078, NHMRC 633000en_AU
dc.subjectMagnetic resonance imagingen_AU
dc.subjectMonte Carlo methodsen_AU
dc.subjectMRI guided radiotherapyen_AU
dc.subjectlung dose enhancementen_AU
dc.titleDose enhancement in radiotherapy of small lung tumors using inline magnetic fields: A Monte Carlo based planning studyen_AU
dc.typeArticleen_AU
dc.subject.asrc029903en_AU
dc.identifier.doi10.1118/1.4938580.
dc.type.pubtypePre-printen_AU


Show simple item record

Associated file/s

Associated collections

Show simple item record

There are no previous versions of the item available.