dc.description.abstractBackground and purpose: Accurate delivery of radiotherapy is critical to achieve optimal treatment outcomes. Interfraction translational IGRT is now standard, and intrafraction motion management is becoming accessible. Some platforms can report both translational and rotational movements in real time. This study aims to quantify the dosimetric impact of observed intrafraction rotation of the prostate measured using monitoring software. Materials and methods: A dose grid resampling algorithm was used to model the dosimetric impact of prostate rotations for 20 patients on a SBRT prostate clinical trial. Translations were corrected before and during treatment, but rotations were not. Real time rotation data were acquired using KIM and a cumulative histogram analysis performed. Prostate volumes were rotated by the range of observed angles and used to calculate DVH data. Results: The pitch axis had a higher range of observed rotations resulting in only 7 patients spending at least 90% of the beam on time across all fractions within rotation angles resulting in PTV D95% ≥36 Gy in this axis. The yaw and roll axes saw 17 and 15 patients respectively achieving this criterion. All but one of 20 patients exceeded CTV D98% ≥36 Gy for all observed rotation angles. Conclusions: Current CTV-PTV margins do not result in compromised CTV dose coverage due to inter and intrafraction prostate rotations in the absence of other uncertainties. Reduced PTV dosing is due to the extremely conformal treatment delivery but is unlikely to be clinically deleterious. Prostate standard IGRT should continue to focus on correcting any observed translational movements. Margin reduction could be explored in conjunction with other uncertainties.en
dc.language.isoenen
dc.publisherElsevieren
dc.relation.ispartofRadiotherapy & Oncologyen
dc.rightsCopyright All Rights Reserveden
dc.subjectDosimetryen
dc.subjectSBRTen
dc.subjectProstateen
dc.titleDosimetric impact of intrafraction rotations in stereotactic prostate radiotherapy: A subset analysis of the TROG 15.01 SPARK trial.en
dc.typeArticleen
dc.subject.asrc0299 Other Physical Sciencesen
dc.identifier.doi10.1016/j.radonc.2019.04.013
dc.relation.nhmrc1112096
dc.relation.other108
usyd.facultySeS faculties schools::Faculty of Medicine and Health::Central Clinical Schoolen
usyd.departmentACRF Image X Instituteen
usyd.citation.volume136en
usyd.citation.spage143en
usyd.citation.epage147en